Infant Intensive Care / Warmer
These units sold new for over 12,000.00 each
********************This unit is over 6 feet tall and weights 250 lbs********************
This unit looks very nice with little or no wear.
1. Bottom Tray - (For storing items, blankets, other items)
2. Bed tray - (Can be tilted (by hand) up or down, sides drop down for easy removal)
3. Mattress - (23.5" x 29.5")
4. X-Ray plate - (under mattress - not pictured)
5. Heater & Light hood unit - (This head can swivel, left or right 65 degrees for xrays)
6. Remote alarm - (Extra Red flashing light, & siren - on top of hood unit)
7. Oxygen gage set w/ one hose
8. (2) Small tray units - (Pictured hanging on sides)
9. Skin temperature sensor - (probe)
10. Oxygen tank holder brackets - (with wrench)
11. Oxygen control unit, mounted to mast - (see below for further details)
The only testing I did is by turning on and feeling the heat from the heater. The heater can be controlled, by the sensor attached to the patient, or manually. I was also able to see the lights activate when I switched from man or skin. Also when sliding the heater switch I could see it add more bars to the bar scale.
*** The temp readings were off, since I didnt have the probe on my skin, it also set the alarm off because of this ..
This unit has a battery backup system, that appears to work (sounds alarms), some flashing buttons.
The temperature is displayed on a red digital display, and the temperature can be set manually.
* I have two of these, the second unit does not have the gauge set. If interested I can send you a link for the photos of the second unit.
Below is the factory specs, the units may or may not have everything stated below, please review the photos for details or email me for further info.
Air-Shields IICS-90 (Infant Intensive Care System)
The warming module head can be moved 65˚ in either direction to facilitate other procedures such as x-ray, without interrupting thermal operation.
All bassinets include a smooth operating mattress tilt.
An organizer panel at the rear of the bassinet supports efficient cable management. Each panel can be opened or removed for cleaning. Each panel is marked to optimize positioning of x-ray cassette.
Takes the audible alarm tone away from the infant area and also displays alarm violations in a remote location promoting a low-noise, stress free environment. The remote alarm will initiate alarm violations at a lower intensity, and will only increase if unattended.
A conveniently located x-ray cassette tray enables x-rays to be taken with minimal disturbance to the patient.
· 2 oxygen inlets and 2 oxygen outlets
· 2 air inlets and 1 air outlet
· 1 vacuum inlet and 1 vacuum outlet
· Flow meter without oxygen take-off
· Pin indexed Oxygen Regulator, yoke, and pressure gauge for “E cylinder with 3ft green DISS hose
· Pin indexed Air Regulator, yoke, and pressure gauge for “E cylinder with 3ft yellow DISS hose
· Tank mounting bracket for two “E size cylinders
Ambient operating temp 68˚F (20˚ C) - to 86˚F (30˚ C)
Altitude Sea level to 1.9 miles (3 km)
Control modes Skin Mode and Manual Mode (0-100% in 25% increments)
Skin mode control range 34.0˚ C - 37.9˚ C
Silence/reset Silences audible alarms for 15 minutes
Manual mode After 10 minutes of operation, an audible “chirp tone (Manual Alert) will sound every 30 seconds. The heater may then be reset for an additional 15 minutes without the heater turning off by pressing the Silence/Reset Switch. If the circuit is not reset, the heater is automatically turned off when 15 minutes of operation have elapsed, the MANUAL Indicator lights steady, and the alarm sounds.
Alarms Probe Fail, Power Fail, Set Point (Skin Mode 1.0˚ C) Remote alarm option Initiates large red lamp for alarm conditions Initiates low volume audible which increases if unattended
Bulb type 50 W Quartz Halogen White Light
Power requirements 100 V AC, 50/60 Hz, 440 W
100V and 110/120V models One 10A
Circuit Breaker (on Power Module of associated equipment)
Circuit Breaker (on Power Module of associated equipment)
Power requirements 100 V AC, 50/60 Hz, 680 W 110/120 V AC, 50/60 Hz, 680 W
Leakage current 100 V, 120 V models: <300 μA 220 V-240 V models: <500 μA
Heating element Quartz (600 W)
Examination light 100 ft. candles at center mattress 1076 lux; 100 lumens per sq. ft.
Skin temperature display Digital LED with 0.1˚ C Resolution
Skin temp. variability 0.2˚ C at temperature equilibrium
Mattress tilt 6˚ - Mattress size in (cm) 29.5 (75) x 23.5 (59.7)
Heater module swivel 65˚
Height Low 75 in (190.5 cm) 76.75 in (195 cm) 71 in (180 cm)
Height High 87 in (221 cm) 79in (201 cm) 77 in (195 cm)
Width, 28 in (71 cm) 28 in (71 cm) 28 in (71 cm)
Depth 49 in (124 cm) 46 in (117 cm) 38.5 in (98 cm)